Background and Purpose: The 100-W holmium laser, in conjunction with a sidefiring 550-µm fiber, can ablate prostate glands as large as 40 g to relieve lower urinary-tract symptoms. We evaluated the effect of various setting combinations on ablation efficiency using beef kidney as a tissue-surrogate model.
Materials and Methods: Beef kidney specimens (mean weight 44.8 ± 3.1 g) were secured in a cylinder, which was submerged in a water-filled tank through which a 27F resectoscope and 550-µm sidefiring fiber were positioned. Four energy/frequency combinations were tested, with each used to treat 10 kidney specimens. The difference between the mean pretreatment and post-treatment weights of each treatment group were compared statistically with Student's t-test.
Results: The largest mean weight difference after treatment (8.94 ± 2.38 g) was achieved using 3.2 J and 25 Hz. This mass reduction was significantly greater than that of all other combinations except 2.5 J and 40 Hz.
Conclusions: Use of the 3.2 J and 25 Hz setting combination resulted in the greatest amount of ablation in this tissue-surrogate model, suggesting that maximal energy settings may provide an advantage in tissue vaporization using the 100-W holmium laser. Clinical assessment must be performed to substantiate these findings.