Abstract
The accuracy of the latest generation Dexcom G7 sensors in individuals with diabetes undergoing hemodialysis has not previously been investigated. Participants with diabetes undergoing hemodialysis were recruited, with paired sensor glucose from Dexcom G7 recorded with plasma glucose analyzed in the laboratory, as well as the Freestyle Precision Pro glucometer and EKF Biosen C-Line analyzer. Ten adults (median age 64.0 [58.0–74.5] years) were recruited. Overall percentage (%) mean and median absolute relative differences were 10.4% and 8.5% for matched laboratory pairs, respectively (n = 720). Diabetes Technology Society error grid analysis showed 99.7%, 100%, and 99.9% of pairs within zones A and B for lab, glucometer, and EKF methods, respectively. This, the first Dexcom G7 accuracy study conducted in people on hemodialysis, demonstrates accuracy and safety when compared with lab reference readings. These data support the accessibility of continuous glucose monitoring (CGM) and hybrid closed-loop systems for people with diabetes on hemodialysis.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.