Abstract
Continuous glucose monitoring (CGM) is proposed as an alternative for glycemic assessment in peritoneal dialysis, but volume overload and anemia may affect sensor accuracy. This is an exploratory analysis of a study of Guardian Connect™ with Guardian Sensor™ 3 in 30 participants with diabetes on continuous ambulatory peritoneal dialysis (CAPD) (age [mean ± standard deviation] 64.7 ± 5.6 years, 23 men, body mass index [BMI] 25.4 ± 3.9 kg/m2, blood hemoglobin [Hb] 10.7 ± 1.3 g/dL). The mean absolute relative difference (MARD) was calculated between paired sensor and YSI 2300 STAT venous glucose readings (n = 941) during an 8-h in-clinic session with glucose challenge. Body composition was evaluated using bioimpedance. The overall MARD was 10.4% (95% confidence interval 9.6–11.7). There were no correlations between BMI, extracellular water, relative hydration index, and lean or fat mass with MARD. No correlations were observed between MARD and Hb (r = 0.016, P > 0.05). In summary, this real-time CGM demonstrated good accuracy in CAPD with minimal influence from body composition and anemia.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.