Abstract
Aim:
To assess the real-world performance of MiniMed™ 780G for Australians with type 1 diabetes (T1D) following advanced hybrid closed loop (AHCL) activation and to evaluate the effect of changing from MiniMed 670/770G to 780G.
Methods:
We analyzed deidentified Carelink™ continuous glucose monitoring (CGM) data from Australian users from January 2020 to December 2022, including the proportion attaining three major consensus targets: Glucose management indicator (GMI <7.0%), time in range (TIR 70–180 mg/dL >70%), and time below range (TBR 70 mg/dL <4%).
Results:
Comparing 670/770G users (n = 5676) for mean ± standard deviation 364 ± 244 days with 780G users (n = 3566) for 146 ± 145 days, the latter achieved a higher TIR (72.6% ± 10.6% vs. 67.3% ± 11.4%; P < 0.001), lower time above range (TAR) (25.5% ± 10.9% vs. 30.6% ± 11.7%; P < 0.001), and lower GMI (6.9% ± 0.4% vs. 7.2% ± 0.4%; P < 0.001) without compromising TBR (1.9% ± 1.8% vs. 2.0% ± 1.8%; P = 0.0015). Of 1051 670/770G users transitioning to 780G, TIR increased (70.0% ± 10.7% to 74.0% ± 10.2%; P < 0.001), TAR decreased (28.1% ± 10.9% to 24.0% ± 10.7%; P < 0.001), and TBR was unchanged. The percentage of users attaining all three CGM targets was higher in 780G users (50.1% vs. 29.5%; P < 0.001). CGM metrics were stable at 12 months post-transition.
Conclusion:
Real-world data from Australia shows that a higher proportion of MiniMed 780G users meet clinical targets for CGM consensus metrics compared to MiniMed 670/770G users and glucose control was sustained over 12 months.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.