Abstract
Objective:
The safety and impact of the advanced hybrid closed-loop (AHCL) system on glycemic outcome in 2- to 6-year-old children with type 1 diabetes and the diabetes distress of caregivers were evaluated.
Research Design and Methods:
This was an open-label prospective study (n = 35) with historical controls matched by treatment unit, diabetes duration, age, gender, and baseline treatment modality. The inclusion criteria were (1) type 1 diabetes diagnosis >6 months, (2) total daily dose of insulin ≥8 U/day, (3) HbA1c <10% (85 mmol/mol), and (4) capability to use insulin pump and continuous glucose monitoring. The MiniMed 780G™ AHCL in SmartGuard™ Mode was used for 12 weeks. Parental diabetes distress was evaluated with a validated Problem Areas In Diabetes—Parent, revised (PAID-PR) survey.
Results:
No events of diabetic ketoacidosis or severe hypoglycemia occurred. Between 0 and 12 weeks, HbA1c (mean change = −2.7 mmol/mol [standard deviation 5.7], P = 0.010), mean sensor glucose value (SG) (−0.8 mmol/L [1.0], P < 0.001), and time above range (TAR) (−8.6% [9.5], P < 0.001) decreased and time in range (TIR) (8.3% [9.3], P < 0.001) increased significantly, whereas no significant change in time below range (TBR) was observed. At the same time, PAID-PR score decreased from 37.5 (18.2) to 27.5 (14.8) (P = 0.006).
Conclusions:
MiniMed 780G™ AHCL is a safe system and 12-week use was associated with improvements in glycemic control in 2- to 6-year-old children with type 1 diabetes. In addition, AHCL is associated with a reduction in parental diabetes distress after 12-week use.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.