Abstract
Background:
Accurate estimation of glycated hemoglobin (HbA1c) from continuous glucose monitoring (CGM) remains challenging in clinic. We propose two statistical models and validate them in real-life conditions against the current standard, glucose management indicator (GMI).
Materials and Methods:
Modeling utilized routinely collected data from patients with type 1 diabetes from central Poland (eligibility criteria: age >1 year, diabetes duration >3 months, and CGM use between 01/01/2015 and 12/31/2019). CGM records were extracted from dedicated Medtronic/Abbott databases and cross-referenced with HbA1c values; 28-day periods preceding HbA1c measurement with >75% of the sensor-active time were analyzed. We developed a mixed linear regression, including glycemic variability indices and patient's ID (glucose variability-based patient specific model, GV-PS) intended for closed-group use and linear regression using patient-specific error of GMI (proportional error-based patient agnostic model, PE-PA) for general use. Models were validated with either new HbA1cs from closed-group patients or separate patient-HbA1c pool. External validation was performed with data from clinical trials. Performance metrics included bias, its 95% confidence interval (95% CI), coefficient of determination (R 2), and root mean square error (RMSE).
Results:
We included 723 HbA1c-CGM pairs from 174 patients (mean age 9.9 ± 4.4 years and diabetes duration 3.7 ± 3.6 years). GMI yielded R 2 = 0.58, with different bias between Medtronic and Abbott devices [0.120% vs. −0.152%, P < 0.0001], and overall 95% CI = −0.9% to +1%, RMSE = 0.47%. GV-PS successfully captured patient-specific variance (closed-group validation: R 2 = 0.83, bias = 0.026%, 95% CI = −0.562% to 0.591%, RMSE = 0.31%). PE-PA performed similarly on new patients (R 2 = 0.76, bias = −0.069%, 95% CI = −0.790% to 0.653%, RMSE = 0.37%). In external validation GMI, GV-PS, and PE-PA produced 73.8%, 87.5%, and 91.0% predictions within 0.5% (5.5 mmol/mol) from the true value.
Conclusion:
Constructed models performed better than GMI. PE-PA provided an accurate estimate of HbA1c with fast and straightforward implementation.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.