Background:
This two-center pilot study combined for the first time an intra-arterial glucose sensor with a decision support system for insulin dosing (SGCplus system) in critically ill patients with hyperglycemia.
Methods:
Twenty-two patients who were equipped with an arterial line and required iv insulin therapy were managed by the SGCplus system during their medical treatment at the intensive care unit.
Results:
Time to target was 111 ± 195 min (80–150 mg/dL) and 135 ± 267 min (100–160 mg/dL) in the lower and higher glucose target group. Mean blood glucose (BG) was 142 ± 32 mg/dL with seven BG values <70 mg/dL. Mean daily insulin dose was 62 ± 38 U and mean daily carbohydrate intake 148 ± 50 g/day (enteral nutrition) and 102 ± 58 g/day (parenteral nutrition). Acceptance of SGCplus suggestions was high (93%).
Conclusions:
The SGCplus system can be safely applied in critically ill patients with hyperglycemia and enables good glycemic control.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.