Abstract
The development of accurate and easy-to-use continuous glucose monitoring (CGM) improved diabetes treatment by providing additional temporal information on glycemia and glucose trends to patient and physician. Although CGM enables users to lower their average glucose level without an increased incidence of hypoglycemia, this comes at the price of additional patient effort. Automation of insulin administration, also known as closed-loop (CL) or artificial pancreas treatment, has the promise to reduce patient effort and improve glycemic control. CGM data serve as the conditional input for insulin automation devices. The first commercial product for partial automation of insulin administration used insulin delivery shutoff at a predefined glucose level. These systems showed a reduction in hypoglycemia. Insulin-only CL devices show increased time spent in euglycemia and a reduction of hypo- and hyperglycemia. Improved glycemic control, coinciding with a minor decrease in hemoglobin A1c level, was confirmed in recent long-term home studies investigating these devices, paving the way for pivotal studies for commercialization of the artificial pancreas. Although the first results from dual-hormone CL systems are promising, because of increased cost of consumables of these systems, long-term head-to-head studies will have to prove superiority over insulin-only approaches. Now CL glucose control for daily use might finally become reality. Improved continuous glucose sensing technology, miniaturization of electrical devices, and development of algorithms were key in making this possible. Clinical adoption challenges, including device usability and reimbursement, need to be addressed. Time will tell for which patient groups CL systems will be reimbursed and whether these devices can deliver the promise that they hold.
Introduction
S
As shown already in 1977, advanced diabetes technology, at that time the Biostator (Ames Division, Miles Laboratories, Elkhart, IN), holds the promise to further alleviate this burden by automation of insulin administration. 4,5 At the time investigators showed that with closed-loop (CL) technology using intravenous glucose sampling and insulin and glucose administration, it was possible to achieve near-normal glucose control in patients with type 1 diabetes mellitus (T1DM). But, due to the complexity, bulkiness, and invasiveness of the procedure, the technology could not be used outside the clinical research center. The development of subcutaneous continuous glucose sensing technology allowed for less invasive glucose sensing, making at-home application of CGM technology a reality. This evolved into combined use of CSII and CGM, also known as the sensor-augmented pump (SAP), and enabled research into automation of insulin administration (artificial pancreas/CL).
Methods and Aims
We aim to provide an overview of the development of CGM technology and its effect on diabetes treatment. We will discuss the impact of CGM on clinical outcomes and innovative approaches and will give an overview of current and ongoing artificial pancreas studies.
We performed a review of published literature on the effectiveness of CGM and CGM use in conjunction with CSII technology in T1DM, using the search terms “CGM,” “CGMS,” and “continuous glucose monitor*.” Studies on diabetes and pregnancy were excluded. Because of the maturity of the field we limited the search results to meta-analyses and systematic reviews. We performed a systematic review of published literature on the topic of CL technology, using the search terms “artificial pancreas,” “closed-loop,” “closed loop,” and “diabetes,” published in the last 5 years. Only outpatient studies with a CL intervention period of at least 4 days performed in patients with T1DM were included. Two hundred seventy-four abstracts were screened; of these, 11 met the inclusion criteria and were included in this review. We asked major CGM manufacturers to provide information about upcoming CGM products and CL technology and performed a review of clinical trial databases, financial reports, and press releases.
CGM Technology and Directions for Future Development
Although different techniques for subcutaneous glucose measurement were introduced, currently only electrochemical transcutaneous CGM systems are available to patients. 6,7 Also, different approaches to access the glucose-containing interstitial fluid, like microdialysis and fully implantable sensors, were tested with varying success. 6,7
Innovative approaches have brought interesting new products, like flash glucose monitoring and implantable CGM systems, to, or close to, market. Flash glucose monitoring (Abbott Diabetes Care, Alameda, CA) represents a new “on-demand” application of subcutaneous glucose sensing technology, effectively using CGM technology for replacement of self-monitoring of blood glucose. 8 Flash glucose monitoring could be seen as CGM on-demand, but the lack of (false-positive) alarms might actually be experienced as a benefit by some. Because of its 2-week longevity, cost per day was reduced, bringing the system within reach for out-of-pocket patient payment and within the reimbursement system in some countries such as Norway and Sweden. Also, factory calibration eliminated the need for frequent recalibration, increasing ease of use.
After failed efforts at the beginning of the century, in the coming year a CGM system using a long-term fully implantable sensor will be brought to the market, first in Europe (Eversense™; Senseonics Inc., Germantown, MD). 9,10 Implantable CGM systems can provide the patient additional ease of use because no sensor is inserted through the skin, and consequently the transmitter can be removed easily and more frequently without the need for sensor replacement. Furthermore, the hassle of weekly sensor replacement with warm-up time and the risk of damage to the inserted sensor is reduced. However, the need for implantation and removal through minor surgery imposes some discomfort on the patient and requires additional effort by the clinician. First conference presentations point toward an accuracy of around that of currently available transcutaneous CGM systems. 9 A new CGM system in development by Roche Diagnostics (Mannheim, Germany), which would be their debut in CGM manufacturing, was reported in 2013 to show an overall mean absolute relative difference of <10% but has not yet reached the market. 11
Improvement in CGM signal filtering levels and calibration algorithms to account for random signal noise and calibration errors resulted in significant improvement of sensor performance. 12 Furthermore, reduction of biofouling and enzyme degradation might reduce signal error and improve reliability and longevity of current CGM devices. 13,14 Variability in initial sensor insertion trauma might account for between sensor variability in CGM performance. 13
Nonetheless, through innovation in sensor technology, the overall accuracy and point precision of CGM systems have improved toward the proposed mark required for making insulin dosing decisions based on CGM data (mean absolute relative difference of <10%). 15 As of September 2015, Dexcom (San Diego, CA) has received approval for nonadjunct use of CGM in the European Union (G5™ Mobile CGM) and appears to be well underway making the same claim with the Food and Drug Administration, providing a statement of confidence in CGM reliability and accuracy by these authorities. This brings legalities in line with daily practice because patients already often use continuous glucose data as a basis for treatment decisions without confirmation of glucose values by self-monitoring of blood glucose (nonadjunct use of CGM).
Currently all CGM devices require initial calibration before displaying CGM data and (re)calibration generally at a 12-h rate during the sensor's lifetime. A further reduction of in vivo sensor-to-sensor differences in sensitivity and sensitivity degradation over the sensor lifetime of current CGM would allow for factory calibration. 16 Factory calibration of CGM eliminates an importance source of user-introduced error and increases the system's ease of use. The ability to use factory-calibrated sensors in CGM-based flash glucose monitoring might indicate that manufacturers are already capable of producing factory-calibrated CGM devices but might prefer to first focus on further improvement of accuracy while maintaining in vivo calibration. 16
Most companies have converted their offline CGM downloads to the “cloud,” allowing for efficient data access by patients and clinicians. Diabetes data management platforms like Glooko® (Glooko, Palo Alto, CA) make data from devices of different manufacturers available to the patient and clinician through a universal portal or smartphone application. 17 Automated CGM data analysis and treatment advice would be a next step for CGM data platforms. Currently no universal CGM training program is available, although the ability to interpret CGM data is important to make effective use of the technology. With what we have learned so far from our patient users, it should be possible to develop such programs. 18
Integrating CGM data in smartphone applications instead of dedicated devices, like with the Dexcom G5 Mobile, might further improve user-friendliness of CGM. Innovation in CGM technology and application of CGM data of main manufacturers are given in Table 1.
Major CGM manufacturers were contacted. We thank Dexcom and Medtronic Diabetes Care for providing these data.
FDA, Food and Drug Administration; SAP, sensor-augmented pump.
Evidence for Clinical Relevance of CGM in T1DM
The JDRF landmark trial, performed in 2008, and its follow-up studies evaluated the benefit of CGM compared with standard glucose monitoring (self-monitoring of blood glucose) for T1DM management in two studies. 19 The study showed that in patients with HbA1c of ≥53 mmol/L (≥7.0%) and above 25 years of age, a reduction of 5.5 mmol/mol (0.5%) in HbA1c was achieved without an increase in hypoglycemia and that wear time was important in reaching this reduction. The relation among wear time of CGM, baseline HbA1c, and HbA1c lowering was later confirmed by a meta-analyses of Pickup et al. 20 A Cochrane meta-analysis 21 showed that CGM technology, mostly studied in conjunction with CSII (SAP), can improve glucose control by means of a reduction in HbA1c level without increase of hypoglycemia. Whether this is true for all patient groups and whether CGM is able to reduce hypoglycemia are still debated. Only 20% of children in the JDRF landmark trial were able to maintain a high wear time, possibly explaining why so far no convincing results have been presented on effectiveness of CGM use in this group.
Although CGM can improve glycemic control and partially relieve the psychological burden of diabetes, 3 patients still need to self-monitor glucose and take carbohydrate content of food and physical activity into account when making treatment decisions. Automation of insulin administration might be able to further alleviate this burden and improve time in the target glucose range. 22
First Steps in Automation of Insulin Administration
The first steps in automation of insulin administration were taken with systems that automate insulin delivery shutoff at very low glucose levels based on CGM data. If the patient is unresponsive to the CGM hypoglycemia alarm, insulin delivery will remain shut off for 2 h or until the patient restarts insulin delivery. Although stopping insulin delivery might be considered only the first form of automation of insulin administration, these systems effectively represented the first approved nonadjunct use of CGM data for insulin administration decisions. From a regulatory perspective, approval of these devices was therefore also important for more advanced forms of automation of insulin administration.
Two randomized controlled trials were performed investigating low glucose suspend (LGS) (Medtronic, Northridge, CA), Bergenstal et al. 23 showed reduced incidence and duration of nonsevere hypoglycemia, whereas Ly et al. 24 showed reduced incidence and duration of nonsevere and severe hypoglycemia compared with the control arm. However, a recent report by the German Institute for Quality and Efficiency in Health Care criticized the results, concluding that presented data were not reliable because of use of events rates and a large baseline difference in prevalence of hypoglycemia in spite of randomization. 25
Recently this “very low glucose, insulin pump off” feature was upgraded with basic predictive technology (predictive LGS [PLGS]) to suspend insulin infusion when the glucose level is expected to fall below 80 mg/dL within 30 min, for a maximum of 2 h or until the glucose level returns to above 70 mg/dL. 26,27 PLGS was also able to reduce the incidence and duration of hypoglycemic episodes (<70/<60 mg/dL), but with an increase in mean overnight glucose level. 26,27 Although an increase in mean glucose level does not necessarily mean worse glycemic control, in the study by Buckingham et al. 26 patients also spent a larger percentage of time in hyperglycemia, indicating that a reduction of hypoglycemia came at the price of hyperglycemia. Maahs et al. 27 showed no increase in hyperglycemia, indicating that the increase in mean glucose level observed in their study is due to a reduction of hypoglycemia only. Currently insulin pump manufacturers other than Medtronic are preparing LGS-like functionalities to be included into their pumps.
Further Automation of insulin administration, the artificial pancreas
CL systems, also known as the “artificial pancreas,” are designed to automate glucose control. A CL system consist of an insulin pump (CSII), a system for real-time glucose sensing (CGM), and algorithms for safety and glucose control. Systems either communicate via wired or wireless connections or are integrated all-in-one devices. 28,29 Insulin-only and dual-hormone approaches using insulin and glucagon or amylin have been investigated. 28,30 –34
Complex algorithms are required to overcome the remaining shortcomings of CGM technology and delay in insulin action, due to both physiology and slow absorption of current “fast-acting” insulin. The most common algorithms are based on model predictive control, proportional integral derivative, or fuzzy logic. 35 Algorithms can be integrated or consist of separate module(s) for safety and glucose regulation. 36 Algorithm self-learning capabilities and integration of auxiliary sensors for detection of exercise might allow for individualized treatment and treatment adaption over time. 37
Various modes of CL operation can be discriminated, ranging from fully automated insulin administration systems (full-CL) requiring virtually no user input to hybrid-CL systems requiring frequent user input, for example, for meal intake or exercise announcement. 30,38 CL systems can be used either day and night or only during a specified period of time like in nighttime-only systems. These systems use a hybrid approach, with automated insulin administration (CL) during one period (e.g., nighttime) and without automated insulin administration for the remainder of the time (e.g., daytime or for meals). After introduction of LGS and PLGS, the hypoglycemia hyperglycemic minimizer could be considered the first-generation artificial pancreas systems. 39
The main results of available artificial pancreas studies are given in Table 2. The first results from longer-term hybrid-CL studies were recently presented. 29,40,44,45 Two studies with a duration of 2–3 months investigating evening and night, as well as day and night, artificial pancreas use showed a reduction in time in hypo- and hyperglycemia and improved mean glucose level over investigated time periods compared with SAP therapy. Both reported a modest reduction in HbA1c level (2–3 mmol/mol, 0.2–0.3%) over SAP. 40 Most of the improvement in glycemia took place during the night period. 40 This seems due to the challenging daytime circumstances with glucose perturbations due to physical activity and carbohydrate intake during meals and snacks. Consequently, depending on the user-friendliness of the first commercialized artificial pancreas systems, artificial pancreas-when-at-home could be considered as a concept for first commercial introduction of the artificial pancreas. 29 Results from these longer-term real-life studies seem to have paved the way for pivotal studies required by the Food and Drug Administration for commercial systems. Current artificial pancreas systems face clinical adoption challenges including improved wearability, connectivity between devices, and user-friendliness of interfaces. Table 3 provides an overview of currently ongoing studies, investigating longer-duration artificial pancreas use and use in specific patient groups like adolescents and children.
Outpatient studies with an intervention period of 4 days or longer were selected.
Outcomes are given as intervention versus control, and over the time period that the closed-loop (CL) system was functioning.
DiAs, Diabetes Assistant; dual-hormone, closed-loop system that uses both insulin and glucagon infusion; HbA1c, hemoglobin A1c; hybrid-CL, mode of closed-loop operation that requires user announcement of meals or other activities to the system; insulin-only, insulin-only closed-loop system; MPC, model predictive control; NA, not available; NS, not significant; PID, proportional integral derivative; SAP, sensor-augmented pump.
Source: ClinicalTrials.gov website (accessed October 7, 2015).
CGM, continuous glucose monitoring; CL, closed-loop; CSII, continuous subcutaneous insulin infusion; DiAs, diabetes Assistant; DKA, diabetic ketoacidosis; dual-hormone, closed-loop system that uses both insulin and glucagon; full-CL, mode of closed-loop operation that requires no user input during operation of the system; hybrid-CL, mode of closed-loop operation that requires user announcement of meals or other activities to the system; insulin-only, insulin only closed-loop system; LGS, low glucose suspend; MPC, model predictive control; PLGS, predictive low glucose suspend; SAP, sensor-augmented pump; T1DM, type 1 diabetes mellitus.
Fully automated CL systems incorporating glucagon administration may provide additional benefits over insulin-only systems and could be considered the third-generation and final stage of CL development. 39 However, there are now convincing data that insulin only–based systems can improve glycemic control, reduce hypo- and hypoglycemia, and improve some aspects of diabetes management burden.
A bifurcation in the artificial pancreas roadmap is therefore proposed in which both insulin-only and dual-hormone artificial pancreases are considered end-stage targets with their own pros and cons. 48 Pros of insulin-only systems might be a slight reduction in HbA1c level compared with SAP, shown so far with insulin-only CL systems (2–3 mmol/mol, 0.2–0.3%). Development of soluble pumpable glucagon, dual-chamber pumps, dual-lumen catheters, and finalization of algorithms could be considered main challenges for the dual-hormone approach, although not all might have to be solved before market introduction. In an inpatient head-to-head study, a dual-hormone artificial pancreas resulted in a significant reduction in the number of hypoglycemic events. 49 In a more recent diabetes camp study a dual-hormone artificial pancreas showed a significant reduction of time spent in hypoglycemia and a trend toward further reduction of hyperglycemia and improved time in euglycemia compared with an insulin-only approach. 50 Longer-term head-to-head studies comparing the best available insulin-only and dual-hormone AP approaches are required for a careful consideration of pros and cons of both systems. Amylin–insulin coformulation could also be considered an option for dual-hormone AP treatment but is not currently being investigated in longer-term CL systems. 34
The added benefit of CL systems over CSII-only might increase patients' willingness for out-of-pocket payment for CGM sensors to upgrade their system with CL functionality, or even for payment by reimbursement authorities. Insulin-only systems could be seen as the natural evolution after SAP and might become available via current CSII + CGM reimbursement programs soon after market introduction. Dual-hormone approaches, because of expected increased cost for device consumables, have to prove added value over SAP and insulin-only systems to be considered for reimbursement. Table 4 provides information on possible commercialization of artificial pancreas technology in the coming years.
CGM, continuous glucose monitoring; DiAs, Diabetes Assistant; dual-hormone, closed-loop system that uses both insulin and glucagon; full-CL, mode of closed-loop operation that requires no user input during operation of the system; hybrid-CL, mode of closed-loop operation that requires user announcement of meals or other activities to the system; insulin-only, insulin-only closed-loop system.
Summary and Outlook
The development of accurate and easy-to-use CGM devices improved diabetes treatment by providing additional temporal information on glycemia and glucose trends to patient and physician. Although CGM enables users to lower their average glucose level without an increased incidence of hypoglycemia, this comes at a price—the price of additional patient effort. Automation of insulin administration has the promise to reduce the need for additional patient effort and improve glycemic control.
In the coming years CGM systems will reach an accuracy level at which further improvement translates into diminished additional clinical benefit, making other features like ease of clinical data application and factory calibration, but also use of CGM for automation of insulin administration, more important. CGM data serve as the input for insulin automation devices, and as such the development of reliable CGM technology was important in the advancement of automated insulin administration. The first commercial product using partial automation of insulin administration used insulin delivery shutoff at a predefined LGS. These systems showed a reduction in (severe) hypoglycemia compared with the most advanced treatment (SAP). Systems using additional basic predictive technology (PLGS) also showed a decrease in hypoglycemia but at the cost of a rise in mean overnight glucose level. Hybrid insulin-only CL devices were the first to show that an increased time spent in euglycemia, and a reduction of hypo- and hyperglycemia can be expected from CL technology. Improved glycemic control, with minor decreases in HbA1c levels, was confirmed in recent long-term home studies investigating these devices, paving the way for pivotal studies for commercialization of AP. Although the first results from dual-hormone CL systems are promising, because of increased cost of consumables of these systems, long-term head-to-head studies will have to prove superiority over insulin-only approaches.
After a long wait since the first experiment with the Biostator, CL glucose control for daily use seems to finally become a reality. Improved continuous glucose sensing technology, miniaturization of electrical devices, and development of algorithms were key in making this possible. University projects have provided the evidence needed for regulatory organizations to allow pivotal trials of commercial devices, while at the same time commercially driven initiatives in basic automation of insulin administration (LGS/PLGS) helped to boost confidence in the technology. Intensive collaborations among funding bodies, commercial entities, and not-for-profit organizations have been essential for the development of CL systems.
Clinical adoption challenges, including device usability and reimbursement, will need to be addressed. This includes improved connectivity between devices and improvements in algorithms and insulin to allow for a switch from hybrid-CL to full-CL products. Time will tell for which patient groups CL systems will be reimbursed and to what extent these devices can deliver the promise they hold.