Abstract
Objective:
This study compared the efficacy, safety, device satisfaction, and quality of life (QOL) in people with diabetes using an insulin bolus-patch versus current devices (pen/syringe) to deliver mealtime insulin.
Research Design and Methods:
Thirty-eight subjects with diabetes (26 with type 1 and 12 with type 2) were randomized to bolus-patch or current injection device (55% pen and 45% syringe) to deliver mealtime insulin in a multicenter, 6-week crossover study. Efficacy was assessed by equivalence in mean daily seven-point blood glucose (MDBG). Safety assessments included severe hypoglycemia episodes, adverse device effects (ADEs), and adverse events (AEs). Device satisfaction was determined by the validated Insulin Delivery System Rating Questionnaire (IDSRQ) and QOL by the validated Diabetes Specific QOL Scale (DSQOLS).
Results:
Using bolus-patch, MDBG (mean±SE) was equivalent to that using pen/syringe (8.61±0.28 vs. 9.02±0.26 mmol/L;
Conclusions:
Delivery of mealtime insulin with bolus-patch compared with pen/syringe resulted in equivalent MDBG, lower SD and CV of seven-point blood glucose measurements, good safety, significant device satisfaction, and improved QOL.
Get full access to this article
View all access options for this article.