Abstract
In vivo subcutaneous glucose sensor accuracy depends on the calibration method. Sensor accuracy was assessed during standard oral glucose tolerance tests in six non-diabetic subjects each wearing six subcutaneous glucose sensors (Medtronic MiniMed). Paired blood glucose (B G) and sensor current readings were used for retrospective sensor calibration using either B G or sensor current as the independent variable. Sensor accuracy after calibration was assessed using three criteria: linear regression between B G and sensor glucose (S G); correlation; and mean absolute difference (MAD), defined as 100 × |BG - S G|/B G. Calibration with B G as the independent variable resulted in unbiased estimates of regression slope (1.02, not different than 1, p< 0.01) and y-intercept (-1.06 mg/dL, not different than 0, p< 0.01). In contrast, calibration with sensor current as the independent variable resulted in biased estimates of slope (0.76, different than 1, p< 0.01) and y-intercept (31.25 mg/dL, different than 0, p< 0.01). However, with sensor current as the independent variable, the MAD was lower than the corresponding value for calibration with B G at the x-axis (15.00 ± 0.47% vs. 18.35 ± 0.63%, p< 0.01). The Pearson correlation coefficient between B G and S G was higher when using sensor current as the independent variable (R = 0.82 vs. R = 0.79 when using glucose on the x-axis). We suggest that despite the fact that calibration with sensor current as the independent variable leads to a bias in the estimate of B G, it is a more appropriate calibration method when the primary concern is minimization of the MAD between S G and B G.
Get full access to this article
View all access options for this article.
