Abstract
Purpose
: The use of stapling devices in pediatric thoracoscopic surgery is limited due to the decreased maneuverability of 12-mm instruments in a child's chest. The goal of this study is to quantify the ability of 5-mm energy devices to seal lung tissue in the nonsurvival swine model.
Methods
: Nine 10-kg female swine were divided into three nonsurvival groups. Group A (n = 3): Left thoracotomy with use of a 12-mm stapler (US Surgical Corporation, Norwalk, Connecticut). Group B (n = 3): Left thoracoscopy with use of the LigaSure LS1000 (Valleylab, Boulder, Colorado) 5-mm instrument. Group C (n = 3): Left thoracoscopy with use of the Ultracision LCS-K5 (Ethicon Endo-Surgery, Cincinnati, Ohio) 5-mm instrument. Biopsy specimens of the lingula of the lung were taken. At the end of the procedure, seal burst pressures were recorded.
Results
: Average burst pressure (mm Hg): A, staples 43.5 (43-44); B, LigaSure 44.9 (40.2-53.6); C, harmonic 37.5 (30-46.4). Average seal length (mm): A, staples 30 (30-30); B, LigaSure 27 (21.4-30); C, harmonic 26 (22-27). Average biopsy weight (g): A, staples 0.52 (0.51-0.53); B, LigaSure 1.78 (1.69-2.14); C, harmonic 1.58 (0.3-1.66). The standard deviations for the pressures were as follows: A, staples 0.5; B, LigaSure 7.54; C, harmonic 8.29. No statistically significant differences between the burst pressures were determined by the t test (A vs. B, P = .78; A vs. C, P = .33). There was 80% power to detect a difference in the means of 25 mm Hg for the LigaSure (A) and of 27 mm Hg for the harmonic (B).
Conclusion
: Both the LigaSure and the harmonic scalpel can effectively seal normal lung tissue from air leaks in the nonsurvival swine model.
Get full access to this article
View all access options for this article.