Abstract
Background and Purpose
: New devices such as the Ti-knot and Hem-o-lok clips have been developed for laparoscopic surgical applications. We compared the effectiveness of Ti-knot TK5 (LSI Solutions), Hem-o-lok MLK clips (Weck Closure), Ligaclip 5-mm titanium clips (Ethicon), and Endopath vascular staples (35 mm long, 12.3 mm wide) (Ethicon).
Materials and Methods
: Renal artery segments from 5 to 6 mm in diameter were harvested from fresh porcine kidneys. One end of the vessel was intubated with a 25-gauge ball-tipped needle and fastened with two silk ties. The other end was occluded with one of the test devices. Saline was infused into each arterial segment at 3 mL/min with the maximum pump pressure at 800 mm Hg. The maximum pressure with leakage was recorded. Each of the five test devices was tested eight times on a rotating basis. Saline infusion was stopped when the maximum pump pressure was reached or when leakage was observed.
Results
: All Ti-knot devices, Hem-o-lok clips, titanium metal clips, and standard hand ties tolerated pressures >800 mm Hg with no leakage, but 4 of the 8 vascular staple lines (50%) leaked before this maximum pump pressure was reached. For those that leaked, the mean leak pressure was 273 mm Hg (range 237-322 mm Hg).
Conclusions
: All devices tested are capable of occluding renal arteries under physiologic conditions. Ti-knot devices and Hem-o-lok clips occluded renal arteries to pressures that exceeded 800 mm Hg. They are equivalent to hand ties under supraphysiologic conditions.