Abstract
Objective
Improvised tourniquets in first aid are recommended when no scientifically designed tourniquet is available. Windlasses for mechanical advantage can be a stick or pencil and can be used singly or multiply in tightening a tourniquet band, but currently there is an absence of empiric knowledge of how well such windlasses work. The purpose of the present study was to determine the performance of improvised tourniquets in their use by the type and number of windlasses to improve tourniquet practice.
Methods
A simulated Leg Tourniquet Trainer was used as a manikin thigh to test the effectiveness of improvised tourniquets of a band-and-windlass design. Two users made 20 tests each with 3 types of windlasses. Tests started with 1 representative of a given type (eg, 1 pencil), then continued with increasing numbers of each windlass type until the user reached 100% effectiveness as determined by cessation of simulated blood flow. Windlass types included chopsticks, pencils, and craft sticks.
Results
Effectiveness percentages in stopping bleeding were associated inversely with breakage percentages. Pulse stoppage percentages were associated inversely with breakage. The windlass turn numbers, time to stop bleeding, the number of windlasses, and the under-tourniquet pressure were associated inversely with breakage. The windlass type was associated with breakage; at 2 windlasses, only chopsticks were without breakage. Of those windlass types that broke, 20.7% were chopsticks, 26.1% were pencils, and 53.2% were craft sticks.
Conclusions
A pair of chopsticks as an improvised tourniquet windlass worked better than pencils or craft sticks.
Introduction
First aid tourniquets have become more commonly used, for example, in the aftermath of the Boston Marathon bombing in 2013. 1 After more than a decade of war, with more veterans returning to the United States with tourniquet experience, and increasing media publicity of civilian use of tourniquets in the United States, the potential benefits to the civilian sector are becoming recognized. To control hemorrhage from compressible extremity wounds, improvised tourniquets are recommended when no commercially designed tourniquet is available. 2 Preliminary evidence from war indicates that improvised tourniquets challenge the user to make them both safe and effective. 2 The common type of improvised tourniquet is of a band-and-windlass design where a band is twisted and tightened by a rodlike windlass. The windlass gives mechanical advantage in tightening the band. Types of windlasses may include sticks and pencils, which can be used singly or multiply. However, evidence of efficacy of improvised tourniquets is based mostly on opinions of clinicians with experience in tourniquets. 2 We were unable to find datasets from which the US Army’s lesson plan for improvised tourniquet use was developed. Although the Army’s lesson plan is detailed in technique steps, and many of the claims seem like common sense, we wished to empirically explore the issue of improvised tourniquet use to identify opportunities for improved care.
There is very little literature available for the layman outdoor enthusiast concerning use of improvised tourniquets in the field. 3 –8 Hence, the purpose of the present study is to provide new knowledge to both the novice and to professionals for use and performance of improvised tourniquets by testing various types and numbers of windlasses.
Methods
This study was conducted under a protocol reviewed and approved by the regulatory office of the US Army Institute of Surgical Research. Tourniquet users were a pair of investigators familiar with military tourniquet training and clinical use of tourniquets. Both investigators were experienced in tourniquet research and the use of the manikin.
The investigators used the HapMed Leg Tourniquet Trainer (CHI Systems, Fort Washington, PA, USA)—a right-thigh body segment (leg number 000F) with an amputation injury just proximal to the knee—in a manner similar to previous reports.9,10 The medial hip–pelvic area had a computer interface that included a smartphone-like touchpad software (version 1.9) internal to the manikin. This device allowed the leg to be used independently and to be operated by user input through finger touch on the pad.
The system reported blood loss volume as calculated using a linear equation from the arterial capacity and number of pulses before hemorrhage control. The touchpad readout of each iteration showed the results, which included effectiveness of bleeding control as evidenced by the above-noted light display, time to hemorrhage control, pressure exerted under the tourniquet, and volume of blood loss. Time to hemorrhage control started when the iteration began, and stopped when the manikin sensed that the thigh was losing no more blood or when the windlass broke. Complete effectiveness (100%) was considered to be cessation of blood loss and the termination of a distal pulse of the affected extremity.
The tourniquet was an improvised tourniquet involving a band-and-windlass design in which the band was a triangular (37 × 37 × 52 inches) cloth made of cotton (muslin; Elwyn Inc, Elwyn, PA, USA). This bandage is a common US military item used for multiple first aid purposes (NSN 6510-00-201-1755). After removing it from its package, the users would partially unfold the olive drab bandage until it was 3.5 inches wide and 52 inches long. Users tightened the tourniquet around the thigh until distal simulated bleeding was believed to have stopped, based on visual inspection of the lights and palpitation for the distal pulse in the device. One windlass turn was a 180° arc, which is the limit of wrist supination in turning the windlass.
The present study was an experiment to determine tourniquet performance as differentiated by type of windlass and the number of that type needed to be effective. We selected windlass types that varied in shape, size, and toughness so as to see potentially varied performance. Windlass types included bamboo chopsticks, pencils, and craft sticks (also known as Popsicle sticks). Tourniquet users tested performance in sets of 20 iterations. Each set of 20 iterations of testing was per user with each type of windlass. Users started with 1 windlass for each type. To determine when effectiveness would reach 100% in a set for each user for each windlass type, users proceeded with each successive set with an increasing number of windlasses until the user eventually reached 100% effectiveness for the set. Once the user reached 100% effectiveness for a type of windlass, the user switched to another windlass type and restarted the set progression. There were 2 tourniquet users.
Statistical analysis included use of descriptive statistics. We used contingency tables to compare variables (eg, breakage [yes/no] by hemorrhage control [yes/no]). A χ2 test was used to assess for differences. Analysis of variance was used to analyze the effects of tourniquet use on factors of interest. Analyses allowed for detection of intervariable associations, eg, if there was an association between the windlass type and hemorrhage control. Logistic regression was used to model the occurrence of breakage (yes/no).
Results
The only windlass type to be effective for all tests with the least number of windlasses was chopsticks, with 2 chopsticks being 100% effective for both users (Table 1). One user was able to make 3 windlasses effective 100% for both pencils and craft sticks, whereas the other user needed 4 windlasses for both pencils and craft sticks to achieve 100% effectiveness.
Effective tests by windlass type, windlass number, and user
NA, not applicable.
When all 3 types of windlasses are considered, effectiveness (absence of pulse and blood flow) was associated inversely with breakage (
Results of windlass breakage by effectiveness in stopping simulated blood flow
Results of windlass breakage by windlass turn number
Data represent sum of all iterations, users, and windlasses.
The breakage percentages of the 2 tourniquet users were not statistically different from each other (
Windlass type was associated with percent breakage (
Results of windlass breakage by windlass type
Data represent sum of all iterations, users, and windlasses.
When the windlass number was pooled for all types of windlass, the windlass number was associated inversely with breakage percentages (
Results of windlass breakage by windlass number
Data represent sum of all iterations, users, and windlasses.
Discussion
The major finding of the present study was that these 3 windlass types performed differently; chopsticks performed better than pencils or craft sticks. Two chopsticks reached 100% effectiveness: chopsticks were the only type to succeed completely at that number of windlasses. The results imply that the mechanical properties of these windlasses differed.
The first minor finding was that the number of windlasses affected performance: if the number was sufficient to avoid breaking, then effectiveness rates were reliably high. If windlasses broke, then effectiveness rates were uniformly dismal. The breakage points were routinely at the point of maximal shear at the windlass-band interface. If hemorrhage had been controlled at the time of breakage, then the time of breakage coincided with the time of loss of hemorrhage control. The proportions of effective iterations by windlass number showed a coherent and expected stepwise rise with each added windlass number.
The second minor finding from our data was that arterial occlusion by improvised tourniquet use was not reliably attained because ineffectiveness percentages were high. This differs from previous experience, which was more favorable after the Boston explosions, 1 but windlasses were not apparently used out-of-hospital in Boston and their tourniquets may have been venous. 2 Tourniquets that close off venous flow are venous tourniquets, and by definition they do not close off arteries and are not arterial tourniquets; venous tourniquets are ineffective in hemorrhage control.
The third minor finding was that turn number was inversely associated with breakage percentage. The association was mostly attributable to the fact that the windlass type and windlass count were confounded in the relationship among breakage, turn number, windlass type, and windlass count. For example, 39 of the 61 breakages after only 1 turn were from using only 1 craft stick.
New hypotheses for further research include screening for better windlasses or deselecting worse windlasses, band types may be screened or deselected, and intersubject variability may be explored further.
Limitations of the present report are related to the narrow scope of the experimental design. The manikin experiment did not look at clinical care complexity but focused on the controlled variables of interest.
Conclusions
In summary, the present study reports a manikin experiment that found that a pair of chopsticks as an improvised tourniquet windlass worked better than pencils or craft sticks.