Abstract
Introduction
The aim of this study was to evaluate the radiation dose and image quality at low tube-voltage pediatric chest computed tomographic angiography (CTA) that applies the same contrast-to-noise ratio (CNR) index as the standard tube voltage technique.
Materials and Methods
Contrast-enhanced chest CTA scans of 100 infants were acquired on a 64-row multidetector computed tomography (MDCT) scanner. In the retrospective study, we evaluated 50 images acquired at 120 kVp; the image noise level was set at 25 Hounsfield units. In the prospective study, we used an 80-kVp protocol; the image noise level was 40 Hounsfield units because the iodine contrast was 1.6 times higher than on 120-kVp scans; the CNR was as in the 120-kVp protocol. We compared the CT number, image noise, CT dose index volume (CTDIvol), and the dose-length product on scans acquired with the 2 protocols. A diagnostic radiologist and a pediatric cardiologist visually evaluated all CTA images.
Results
The mean CTDIvol and the mean dose-length product were 0.5 mGy and 7.8 mGy-cm for 80- and 1.2 mGy and 20.8 mGy-cm for 120-kVp scans, respectively (
Conclusions
With the CNR index being the same at 80-kVp and 120-kVp imaging, the radiation dose delivered to infants subjected to chest CTA can be reduced without degradation of the image quality.